ST-766SS Series Oxidizer, pH, ORP & Conductivity Sensors

ST-766SS Series Oxidizer, pH, ORP & Conductivity Sensors

The Pyxis ST-766SS is one stainless steel inline sensor that measures a disinfectant together with pH, ORP, conductivity and temperature. Each model is set up for one oxidiser (or, in the sulfite models, a reducing agent), and uses the pH and temperature it measures to correct that reading automatically.

There are no membranes, no electrolyte gel and no reagents. The sensor reads continuously, needs no burn-in period, and sends its results through two 4–20 mA outputs and RS-485 Modbus. Set-up and calibration can be done wirelessly through the uPyxis® app with an MA-CR or PowerPACK Bluetooth® adapter.

Models

  • Chlorine: FCL (free), TCL (total), TFCL (free and total together), NCL (monochloramine)
  • CLO (chlorine dioxide), Br (bromine), O3 (ozone)
  • H2O2 (hydrogen peroxide), PAA (peracetic acid), DBNPA
  • SO3 (sulfite), and DCL (free chlorine and sulfite, for dechlorination)

Replaceable Electrode Head

The measuring electrodes sit in a front-loading EH-766 head with a flat face that resists fouling and wipes clean. When it wears out, typically after about two years, the head is swapped on site and the sensor body stays in service.

ST-766SS Series Specifications

ItemST-766SS Series
pH / Temperature CompensationAutomatic, up to pH 9.0
Lower Detection LimitAs low as 0.001 ppm (adjustable sensitivity)
pH Range / Precision0.00 – 14.00 pH; ±0.01 pH
ORP Range / Precision-1,500 to +1,500 mV; ±1.0 mV
Conductivity Range / Precision0 – 10,000 µS/cm; ±5% of reading
TemperatureBuilt-in RTD
MeasurementContinuous
Response TimeT95 240 s for the oxidiser; T95 5 s for pH, ORP and conductivity
Sample Temperature4 – 49 °C (40 – 120 °F)
Sample Inlet Pressure0.05 – 0.6 MPa (7.25 – 87 psi)
OutputsTwo 4–20 mA and RS-485 Modbus, 8-pin connector
Wireless AccessBluetooth® 5.0 via MA-CR or PowerPACK adapter, uPyxis® app
Power Supply22 – 26 V DC, about 2 W
Sensor Body316L stainless steel
ProtectionIP67
SelectivityNon-selective; cross-sensitive to other oxidising species
Reference MethodsUS EPA 334.0 / ISO 7393
ApprovalsCE, RoHS, UKCA
Mounting (sold separately)Clean water: FR-50 flow reservoir, ST-007 or ST-009 stainless steel tee. Dirty water: FR-306 PLUS self-brushing flow reservoir
Cables IncludedMA-1.5CR 1.5 m flying lead cable; MA-4.9CR 1.5 m bulkhead cable
Typical Electrode Life2 years
WarrantyElectrode 6 months; sensor body 13 months

ST-766SS Models

ST-766SS ModelMeasuresRangePrecisionElectrode Head4–20 mA OutputsPart Number
FCLFree chlorine0.00 – 5.00 ppm±0.01 mg/L or ±1%EH-766Free chlorine, pH51154
CLOChlorine dioxide0.00 – 5.00 ppm±0.01 mg/L or ±1%EH-766Chlorine dioxide, pH55821
TCLTotal chlorine0.00 – 5.00 ppm±0.01 mg/L or ±1%EH-766Total chlorine, pH51379
NCLMonochloramine0.00 – 5.00 ppm±0.01 mg/L or ±1%EH-766Monochloramine, pH54731
TFCLFree + total chlorine0.00 – 5.00 ppm±0.01 mg/L or ±1%EH-766-01Total chlorine, free chlorine55659
O3Ozone0.00 – 2.00 ppm±0.01 mg/L or ±1%EH-766-03Ozone, pH54828
BrBromine0.00 – 5.00 ppm±0.01 mg/L or ±1%EH-766Bromine, pH52743
H2O2Hydrogen peroxide0.00 – 200.00 ppm±0.5 mg/L or ±1%EH-766Hydrogen peroxide, pH53302
PAAPeracetic acid0.00 – 100.00 ppm±0.5 mg/L or ±1%EH-766-01Peracetic acid, pH54803
SO3Sulfite0 – 100 ppm±0.5 mg/L or ±1%EH-766Sulfite, pH58305
DCLFree chlorine + sulfiteFree chlorine 0.00 – 5.00 ppm; sulfite 0 – 100 ppmFree chlorine ±0.01 mg/L or ±1%; sulfite ±0.5 mg/L or ±1%EH-766-01Free chlorine, sulfite57300
DBNPADBNPA0 – 20 ppm±0.05 mg/L or ±1%EH-766DBNPA, pH50484

Ideal for Cooling, Process & Drinking Water

Wherever a disinfectant has to be held at the right level, the ST-766SS shows the dose and the water conditions that affect it in one reading. It is used in cooling water, drinking water, RO feedwater, wastewater effluent, irrigation and industrial process water, and in food and beverage plants for wash water, flumes, conveyors and equipment sanitising. Clean water installs in an FR-50 flow reservoir or ST-007/ST-009 stainless steel tee; dirty water uses the self-brushing FR-306 PLUS.

Key Features

  • Oxidiser plus pH, ORP, conductivity and temperature from one sensor
  • Twelve models covering chlorine, chlorine dioxide, bromine, ozone, peroxide, peracetic acid, DBNPA and sulfite
  • Automatic pH and temperature compensation of the oxidiser reading, up to pH 9.0
  • No membranes, gels or reagents; no burn-in time
  • Adjustable sensitivity, with a lower detection limit as low as 0.001 ppm
  • Replaceable EH-766 electrode head, typical life about two years
  • Flat front face that resists fouling and is easy to clean
  • Two 4–20 mA outputs and RS-485 Modbus
  • Bluetooth® set-up and calibration through the uPyxis® app
  • 316L stainless steel body, IP67