The Pyxis UC-80 is a touch screen display and data logger for up to four Pyxis inline sensors, connected over RS-485 Modbus. It shows live readings and trend charts for each sensor, provides the calibration screens, and records every input for download over USB.
It can work as a stand-alone station or send its readings to a controller, PLC or DCS through 4–20 mA outputs and Modbus RTU or Modbus TCP. Pyxis sets each unit up for the sensors it will be used with before it ships, so it is ready to run when it arrives.
- UC-80: 4.3″ touch screen; two 4–20 mA inputs and two 4–20 mA outputs
- UC-80 Plus: larger 7″ touch screen; four 4–20 mA inputs, eight 4–20 mA outputs and LoRa™ wireless
Unlike the compact 24 V DC UC-50A, which serves one sensor with push buttons, the UC-80 takes up to four sensors on a touch screen with the 4G CloudLink™ built into every unit, while the UC-100A offers a 7″ screen and Ethernet with 4G as an option.
Built-In 4G Cloud Connection
Every UC-80 and UC-80 Plus contains the Pyxis 4G LTE CloudLink™ gateway and a global SIM card. Once Pyxis activates it (on request, after enrolment), the readings can be watched live and trended in the mobile app, downloaded and reported on from anywhere.
UC-80 Series Specifications
| Item | UC-80 | UC-80 Plus |
|---|---|---|
| Part Number | 14003 | 72875 |
| Display | 4.3" colour industrial capacitive touch screen (LCD) | 7" colour industrial capacitive touch screen (LCD) |
| Sensor Connection | Up to 4 Pyxis sensors via RS-485 Modbus; preconfigured by Pyxis | |
| Outputs | 2 × 4–20 mA; 2 × RS-485 Modbus RTU; Modbus TCP | 8 × 4–20 mA; 2 × RS-485 Modbus RTU; Modbus TCP |
| Inputs | 2 × 4–20 mA; 2 × RS-485 Modbus RTU | 4 × 4–20 mA; 2 × RS-485 Modbus RTU |
| Relays | 2 × 24 V DC relays, passive or active output (user selected) | |
| LoRa™ | Not included | Included |
| Pyxis 4G LTE CloudLink™ | Built in with global SIM card; activated on request after enrolment with Pyxis | |
| Measurement Interval | Continuous | |
| Data Storage | 128 MB; up to 1 million data and event records | |
| USB | 1 × USB host for data download and screen upgrades | |
| Power Supply | 96 – 260 V AC, 50 – 60 Hz, 60 W | |
| Dimensions (H × W × D) | 250 × 150 × 100 mm | 300 × 200 × 190 mm |
| Weight (approx.) | 5 kg | 7 kg |
| Operating Temperature | 4 – 45 °C (40 – 113 °F) | |
| Storage Temperature | −20 – 55 °C (−4 – 131 °F) | |
| Sample Water Temperature | 4 – 40 °C (40 – 104 °F) | |
| Relative Humidity | 20 – 90%, non-condensing | |
| Altitude | Below 2,000 m (6,561 ft) | |
| Protection | IP64 | |
| Approvals | CE, RoHS | |
Ideal for Remote Monitoring & Data Logging
The UC-80 suits sites where several Pyxis sensors need one screen and the data needs to be seen away from the plant: a cooling or process water system, a treatment plant or an unattended site. It logs everything locally and, through the 4G link, puts the same readings on a phone or computer, so the system can be checked without a site visit.
Key Features
- 4.3″ colour capacitive touch screen (7″ on the UC-80 Plus)
- Up to four Pyxis sensors over RS-485 Modbus
- Live readings and trend charts for every input
- Sensor calibration from the screen
- 4–20 mA outputs, Modbus RTU and Modbus TCP
- Two 24 V DC relays, passive or active output
- Up to one million readings and events stored; USB download
- Built-in Pyxis 4G LTE CloudLink™ with global SIM card
- Preconfigured by Pyxis for your sensors
- 96–260 V AC supply, IP64 enclosure