Chemtrol Australia Product - LT-630 Series Self-Wiping Turbidity Sensors

Looking for a quote or to discuss your needs?

You can also review the documentation for technical details and purchasing information.

LT-630 Series Self-Wiping Turbidity Sensors

LT-630 Series Self-Wiping Turbidity Sensors

Turbidity in Dirty Water, up to 4,000 NTU

The Pyxis LT-630 series are submersible turbidity sensors that keep their own optics clean. A wiper sweeps the sensor face on a timer, so air bubbles, slime and settled solids do not build up and drag the reading off. They are made for raw, storm and process water and for discharges, where turbidity is high and the water is far from clean.

The transmitter is built in. Each sensor gives isolated 4–20 mA and RS-485 Modbus outputs for a controller, PLC or DCS, and can be set up, calibrated and checked over Bluetooth® with the uPyxis® app.

  • LT-631: 0 – 500 NTU
  • LT-632: 0 – 1,000 NTU
  • LT-633: 0 – 4,000 NTU

Measuring clean water? For drinking water, filtered water and anything below 1 NTU, the LT-73X ultra-low turbidity sensors read down to 0.001 NTU. The LT-630 covers the other end of the scale: submersed in tanks, channels and basins, with a wiper to cope with dirty water.

Self-Wiping Optics

Two infrared light sources measure the light scattered by particles in the water, in line with ISO 7027. A 316L stainless steel arm with a PTFE blade wipes the lens at an interval you set in the uPyxis app, or on demand. Kept clean, the sensor holds its calibration within 0.1 NTU for up to a year. It can be calibrated with formazin standards or with a sample of your own water whose turbidity has been measured.

LT-630 Series Specifications

ItemLT-631LT-632LT-633
Part Number532495325010002
ParameterTurbidity
Measurement Range0.00 – 500.00 NTU0.00 – 1,000.00 NTU0.00 – 4,000.00 NTU
4–20 mA Output Scale0 – 500 NTU0 – 1,000 NTU0 – 4,000 NTU
Unit of MeasureNTU (nephelometric turbidity unit)
Lower Detection Limit0.5 NTU
Resolution0.1 NTU
PrecisionBelow 100 NTU: ±2 NTU or ±2% of reading; above 100 NTU: ±10% of reading
Accuracy±2 NTU or ±3% of reading
StabilityWithin 0.1 NTU for up to 1 year without calibration, when kept clean
Response TimeT95 < 12 seconds
Measuring MethodScattered light, dual infrared light sources, ISO 7027 compliant
CalibrationFormazin calibration standard or field-tested process sample
Calibration Points (uPyxis app)3 (zero, mid, high)3 (zero, mid, high)5
Self-CleaningWiper arm with PTFE blade; wiping interval set in the uPyxis app
Power Supply22 – 26 V DC, 6 W (supply rated for 250 mA)
OutputsIsolated 4–20 mA (passive, not loop powered) and isolated RS-485 Modbus RTU
Operating Temperature0 – 50 °C (32 – 122 °F)
Storage Temperature−7 – 60 °C (20 – 140 °F)
Maximum Operating Pressure3.1 bar (45 psi)
InstallationSubmersed: fixed on ¾" MNPT conduit, or hung by lifting ring on chain or cableSubmersed (fixed on ¾" MNPT conduit, or hung on chain or cable), or inline in the ST-008 teeSubmersed (fixed on ¾" MNPT conduit, or hung on chain or cable), or inline in the ST-008 tee
Minimum Clearance (submersed)15 cm from vessel bottom; 10 cm from vessel walls
Materials316L stainless steel body and wiper arm; PTFE wiper blade
ProtectionIP68
Dimensions (L × W × H)254 × 57 × 57 mm
Weight1,530 g (3.3 lb)
Cable10 m (33 ft) cable with IP67 8-pin connector, plus 1.5 m (4.9 ft) flying lead with IP67 8-pin connector
Supplied WithMA-1.5CR flying lead, MA-CR Bluetooth® adapter, MA-NEB USB adapter
ApprovalsCE, RoHS
Warranty13 months (Pyxis standard warranty)

Ideal for Raw, Storm and Process Water

The LT-630 series suits raw and surface water intakes, stormwater, process water and treated discharges, where turbidity runs high and sensors foul quickly. It can sit fixed in a tank or channel on ¾” conduit, hang from its lifting ring on a chain or cable (for example to profile the sludge bed in a clarifier or thickener), or, for the LT-632 and LT-633, run inline in a bypass line through the ST-008 tee.

Key Features

  • Three ranges: 0–500, 0–1,000 and 0–4,000 NTU
  • Automatic wiper keeps the lens clear of bubbles and deposits
  • Wiping interval set in the uPyxis® app, with a manual wipe on demand
  • Dual infrared light sources, ISO 7027 compliant
  • Resolution 0.1 NTU; response T95 under 12 seconds
  • Built-in transmitter: isolated 4–20 mA and RS-485 Modbus RTU
  • Bluetooth® or USB set-up, calibration and diagnostics with the MA-CR adapter
  • Calibration holds within 0.1 NTU for up to a year when kept clean
  • 316L stainless steel body, IP68, 24 V DC at 6 W
  • Fixed on conduit with the MA-120-B bracket, hung on a chain, or inline in the ST-008 tee