IK-739-PLUS & IK-736-PLUS Auto-Brushing Turbidity Analyzers
- Product Details
- Specifications
- Drinking & Process Water
- Installation
IK-739-PLUS & IK-736-PLUS Auto-Brushing Turbidity Analyzers
The Pyxis IK-73X-PLUS analysers measure turbidity continuously, with a resolution of 0.002 NTU. Each one comes complete on a single wall panel, set up, tested and calibrated at the factory. Connect the sample line, the drain and the power, and it is ready to run.
- IK-739-PLUS: 0.000 – 40.00 NTU, with the LT-739 sensor
- IK-736-PLUS: 0.000 – 1,000 NTU, with the LT-736 sensor
The panel holds the turbidity sensor in its brushing flow cell, an FS-100 ultrasonic flow meter with a motorised valve that keeps the sample flow at a set rate, and a 7-inch touch screen that displays, logs and passes on the readings.
A Brush Keeps the Lens Clean
The sensor looks through a flat quartz window. A motorised brush in the FT-100-PLUS flow cell wipes it as often as you set, so fouling can’t build up, and the flat window stops air bubbles settling on it. The reading stays steady with very little attention.
Specifications
| Item | IK-739-PLUS | IK-736-PLUS |
|---|---|---|
| Part Number | 12001 | 12003 |
| Turbidity Range | 0.000 – 40.00 NTU | 0.000 – 1,000 NTU |
| Sensor | LT-739 | LT-736 |
| Light Source | Warm white LED, 90° surface scatter | |
| Method | USEPA 180.1 (US) | |
| Accuracy | 0.001 NTU or ±1% of full scale | |
| Resolution | 0.002 NTU | |
| Response Time | 4 seconds | |
| Measurement | Continuous | |
| Flow Cell | FT-100-PLUS with motorised brush | |
| Display | 7-inch colour capacitive touch screen | |
| Data Storage | 4 GB, up to 1 million readings and events | |
| Outputs | Two 4–20 mA, RS-485 Modbus RTU, Modbus TCP | |
| Inputs | Two 4–20 mA, RS-485 Modbus RTU | |
| USB | One USB host port, for data download and screen updates | |
| Network | RJ-45 Ethernet, Modbus TCP | |
| Power Supply | 96 – 260 V AC, 50/60 Hz, 200 W, 10 A fuse | |
| Sample Pressure | 0.05 – 0.67 MPa (7.25 – 100 psi) | |
| Sample Flow | 100 – 2,000 mL/min | |
| Sample Temperature | 4 – 40 °C (40 – 104 °F) | |
| Panel Operating Temperature | 4 – 45 °C (40 – 113 °F) | |
| Storage Temperature | Instrument −20 – 55 °C (−4 – 131 °F); sensors 0 – 50 °C (32 – 122 °F) | |
| Sample Connections | ½" NPT | |
| Flow Meter | FS-100 ultrasonic (transit time), 0 – 3,000 mL/min, 1 mL/min or ±2% of reading, 1.44" colour display | |
| Protection | IP65 panel display, IP67 sensors | |
| Relative Humidity | 20 – 90%, non-condensing | |
| Altitude | Up to 2,000 m (6,561 ft) | |
| Dimensions | 750 H × 450 W × 243 D mm | |
| Weight | About 15 kg | |
| Approvals | CE, RoHS | |
Spare Parts and Calibration
| Part | Part Number |
|---|---|
| FT-100-PLUS auto-brushing flow cell | 16005 |
| FTP-100-1 brush and seal kit for FT-100-PLUS | 28698 |
| LT-739 turbidity sensor, 0 – 40 NTU | 53221 |
| LT-736 turbidity sensor, 0 – 1,000 NTU | 53215 |
| FS-100 ultrasonic flow meter with display | 54200 |
| UC-100A display and data logger | 43054 |
| L-CAL portable calibration kit | 53247 |
| Turbidity standard 10 NTU, 500 mL | 57010-4 |
| Turbidity standard 30 NTU, 500 mL | 57010-8 |
| Turbidity standard 500 NTU, 500 mL | 57010-2 |
The IK-73X-PLUS is for sites where turbidity has to be watched around the clock: drinking water and secondary supply, raw surface water, cooling and process water, and wastewater effluent. It runs on a mains plug, logs every reading on the panel, and sends the results on to a PLC or SCADA system.
Key Features
- Resolution 0.002 NTU; range 0 – 40 NTU or 0 – 1,000 NTU
- Warm white LED at 90° scatter, to USEPA 180.1 (US)
- Motorised brush cleans the sensor window on a programmable cycle
- Flat quartz window resists air-bubble interference
- FS-100 ultrasonic flow meter and motorised valve hold the sample flow steady, with a no-flow alarm
- 7-inch colour touch screen with calibration and diagnostics built in
- Stores up to 1 million readings and events, downloaded by USB
- Two 4–20 mA outputs, RS-485 Modbus RTU and Modbus TCP over Ethernet; two 4–20 mA inputs
- Set up, tested and calibrated at the factory
- Calibrated on site with the L-CAL portable calibration kit
- Mounting: wall panel 750 H × 450 W × 243 D mm, about 15 kg. Use fixings rated for the weight.
- Power: plugs into a standard mains outlet.
- Sample: inlet pressure 0.05 – 0.67 MPa (7.25 – 100 psi), flow 100 – 2,000 mL/min, water 4 – 40 °C.
- Connections: ½” NPT inlet and outlet. The outlet must drain by gravity.
- Upkeep: clean the sensor and flow cell monthly or as needed, replace the brush head yearly, and calibrate every six months (the low point only, for clean water).